Capturing Other Waveforms to the Arbitrary Waveform

The Edit Waveform Menu is opened by pressing Edit Waveform in the main
Waveform Generator Menu.
To capture another waveform to the arbitrary waveform:
1 Press Source; then, use the Entry knob to select the analog channel,
2 Press Store Source to Arb.
When you create a new arbitrary waveform, the existing arbitrary waveform definition
C A U T I O N
is overwritten. Note that you can use the [Save/Recall] key and menu to save
arbitrary waveforms to one of four internal storage locations or to a USB storage
device, and you can recall them later. See
and
If the source waveform frequency and/or voltage exceed the capabilities of the waveform
N O T E
generator, the arbitrary waveform will be limited to the capabilities of the waveform
generator. For example, a 20 MHz waveform captured as the arbitrary waveform, becomes a
12 MHz waveform.

To output the waveform generator sync pulse

1 If the Waveform Generator Menu is not currently displayed on the
2 In the Waveform Generator Menu, press the Settings softkey.
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
All points are adjusted to maintain uniform time spacing between
points.
math, or reference location whose waveform you wish to capture.
"To recall arbitrary waveforms"
The source waveform is decimated into 8192 (maximum) or fewer
arbitrary waveform points.
oscilloscope's softkeys, press the [Wave Gen] key.
"To save arbitrary waveforms"
on page 320.
Waveform Generator
on page 316
17