22
Reference
X Units
Y Units
Date
Time
Frame
Waveform Label
Time Tags
Segment Index
Waveform Data
A waveform may have more than one data set. Each waveform data set
Header
will have a waveform data header. The waveform data header consists of
information about the waveform data set. This header is stored
immediately before the data set.
Waveform Data
Header Size
Buffer Type
380
A 32-bit integer that identifies the unit of measure for X values in the acquired
data:
• 0 = Unknown.
• 1 = Volts.
• 2 = Seconds.
• 3 = Constant.
• 4 = Amps.
• 5 = dB.
• 6 = Hz.
A 32-bit integer that identifies the unit of measure for Y values in the acquired
data. The possible values are listed above under X Units.
A 16-byte character array, left blank in InfiniiVision oscilloscopes.
A 16-byte character array, left blank in the InfiniiVision oscilloscopes.
A 24 byte character array that is the model number and serial number of the
oscilloscope in the format of: MODEL#:SERIAL#.
A 16 byte character array that contains the label assigned to the waveform.
A 64-bit double, only used when saving multiple segments (requires
segmented memory option). This is the time (in seconds) since the first
trigger.
A 32-bit unsigned integer. This is the segment number. Only used when saving
multiple segments.
A 32-bit integer that is the size of the waveform data header.
A 16-bit short that is the type of waveform data stored in the file:
• 0 = Unknown data.
• 1 = Normal 32-bit float data.
• 2 = Maximum float data.
• 3 = Minimum float data.
• 4 = Not used in InfiniiVision oscilloscopes.
• 5 = Not used in InfiniiVision oscilloscopes.
• 6 = Digital unsigned 8-bit char data (for digital channels).
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide