4 Press the Display On softkey to enable the measurement statistics display.
268
The source channel of the measurement is shown in parenthesis after
the measurement name. For example: "Freq(1)" indicates a frequency
measurement on channel 1.
The following statistics are shown: Name of the measurement, current
measured value, mean, minimum measured value, maximum measured
value, standard deviation, and the number of times the measurement
has been made (count). Statistics are based on the total number of
captured waveforms (count).
The standard deviation shown in Statistics is calculated using the same
formula used for calculating the standard deviation measurement. The
formula is shown in the section titled
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
"AC RMS"
on page 253.