CAN/LIN Triggering and Serial Decode
2 In the Trigger Menu, press the Trigger softkey; then, turn the Entry knob
3 Press the Trigger: softkey; then, turn the Entry knob to select the trigger
398
to select the serial slot (Serial 1 or Serial 2) on which the CAN signal
is being decoded.
condition:
Sync (Sync Break) — The oscilloscope triggers on the rising edge at
the Sync Break exit of the LIN single- wire bus signal that marks the
beginning the message frame.
ID (Frame ID) — The oscilloscope triggers when a frame with an ID
equal to the selected value is detected. Use the Entry knob to select
the value for the Frame ID.
ID & Data (Frame ID and Data) — The oscilloscope triggers when a
frame with an ID and data equal to the selected values is detected.
When triggering on a frame ID and data:
To select the frame ID value, press the Frame ID softkey, and use
the Entry knob.
Note that you can enter a "don't care" value for the frame ID and
trigger on data values only.
To set up the number of data bytes and enter their values (in
hexadecimal or binary), press the Bits softkey to open the LIN Bits
Menu.
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide