I2S Triggering and Serial Decode
6 Press the Base softkey and select a number base for entering data
440
If you select an "Armed" value that is equal to or greater than the
"Trigger" value, the "Trigger" value will be increased so that it is
always greater than the "Armed" value.
Decreasing value — similar to the description above except the trigger
occurs on a decreasing data word value, and the "Armed" value is
the value to which the data must rise in order to re- arm the trigger.
values:
Binary (2's complement).
When Binary is selected, the Bits softkey appears. This softkey opens
the I2S Bits Menu for entering data values.
When the trigger qualifier requires a pair of values (as with In
Range, Out of Range, Increasing value, or Decreasing value), the first
softkey in the I2S Bits Menu lets you select which value of the pair.
In the I2S Bits Menu, press the Bit softkey and rotate the Entry knob
to select each bit; then, use the 0 1 X softkey to set each bit value to
zero, one, or don't care. You can use the Set all Bits softkey to set all
bits to the value chosen on the 0 1 X softkey. Don't care values are
only allowed with Equal or Not Equal trigger qualifiers.
Signed decimal.
When Decimal is selected, the softkey(s) to the right let you enter
decimal values with the Entry knob. These softkeys can be Data, <, >,
or Threshold depending on the selected trigger qualifier.
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide