In the serial decode display, the most significant bit is always displayed on the left
N O T E
regardless of how Bit Order is set.

UART/RS232 Triggering

To set up the oscilloscope to capture UART/RS- 232 signals, see
UART/RS232 Signals"
To trigger on a UART (Universal Asynchronous Receiver/Transmitter)
signal connect the oscilloscope to the Rx and Tx lines and set up a trigger
condition. RS232 (Recommended Standard 232) is one example of a UART
protocol.
1 Press [Trigger].
2 In the Trigger Menu, press the Trigger softkey; then, turn the Entry knob
3 Press the Trigger Setup softkey to open the UART/RS232 Trigger Setup
4 Press the Base softkey to select Hex or ASCII as the base displayed on
5 Press the Trigger softkey and set up the desired trigger condition:
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
to select the serial slot (Serial 1 or Serial 2) on which the UART/RS232
signals are being decoded.
Menu.
the Data softkey in the UART/RS232 Trigger Setup Menu.
Note that the setting of this softkey does not affect the selected base of
the decode display.
Rx Start Bit — The oscilloscope triggers when a start bit occurs on Rx.