• I = integrated waveform.
• y = channel 1, 2, 3, 4, or Math 1, Math 2, Math 3 (lower math function)
• co = arbitrary constant.
• i = data point index.
The integrate operator provides an Offset softkey that lets you enter a DC
offset correction factor for the input signal. Small DC offset in the
integrate function input (or even small oscilloscope calibration errors) can
cause the integrate function output to "ramp" up or down. This DC offset
correction lets you level the integrate waveform.
0 V level
Integrate
without
DC offset
correction
Figure 8
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
t = point- to- point time difference.
data points.
Integrate Without Signal Offset