10 Press the
11 Press the Sample Point softkey; then, turn the Entry knob to select the
12 Press the Standard softkey; then, turn the Entry knob to select the LIN
13 Press the Sync Break softkey and select the minimum number of clocks

LIN Triggering

To set up the oscilloscope to capture a LIN signal, see
Signals"
LIN triggering can trigger on the rising edge at the Sync Break exit of the
LIN single- wire bus signal (that marks the beginning of the message
frame), the Frame ID, or the Frame ID and Data.
A LIN signal message frame is shown below:
1 Press [Trigger].
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
Back/Up key to return to the LIN Signals Menu.
sample point at which the oscilloscope will sample the bit value.
One Bit
60%
70%
Sample Point
80%
standard you are measuring (LIN 1.3 or LIN 2.0).
For LIN 1.2 signals, use the LIN 1.3 setting. The LIN 1.3 setting assumes
the signal follows the "Table of Valid ID Values" as shown in section A.2
of the LIN Specification dated December 12, 2002. If your signal does
not comply with the table, use the LIN 2.0 setting.
that define a sync break in your LIN signal.
on page 396.
Sync
Sync
Identifier
Break
Field
Break
Sync Break Exit
CAN/LIN Triggering and Serial Decode
Data
Checksum
Fields
Field
23
"Setup for LIN
397