Interpreting Uart/Rs232 Lister Data; Searching For Uart/Rs232 Data In The Lister - Agilent Technologies 4000 X-Series User Manual

Infiniivision 4000 x-series
Hide thumbs Also See for 4000 X-Series:
Table of Contents

Advertisement

Interpreting UART/RS232 Lister Data

In addition to the standard Time column, the UART/RS232 Lister contains
these columns:
• Rx — receive data.
• Tx — transmit data.
• Errors — highlighted in red, Parity Error or Unknown Error.
Aliased data is highlighted in pink. When this happens, decrease the
horizontal time/div setting and run again.

Searching for UART/RS232 Data in the Lister

The oscilloscope's search capability lets you search for (and mark) certain
types of UART/RS232 data in the Lister. You can use the [Navigate] key
and controls to navigate through the marked rows.
1 With UART/RS232 selected as the serial decode mode, press [Search].