Acquiring to Segmented Memory

When capturing multiple infrequent trigger events it is advantageous to
divide the oscilloscope's memory into segments. This lets you capture
signal activity without capturing long periods of signal inactivity.
Each segment is complete with all analog channel, digital channel (on MSO
models), and serial decode data.
When using segmented memory, use the Analyze Segments feature (see
"Measurements, Statistics, and Infinite Persistence with Segmented
Memory"
segments. See also
To acquire to
1 Set up a trigger condition. (See
segmented
memory
2 Press the [Acquire] key in the Waveform section of the front panel.
3 Press the Segmented softkey.
4 In the Segmented Memory Menu, press the Segmented softkey to enable
5 Press the # of Segs softkey and turn the Entry knob to select the
6 Press the [Run] or [Single] key.
The oscilloscope runs and fills a memory segment for each trigger event.
When the oscilloscope is busy acquiring multiple segments, the progress is
displayed on screen. The oscilloscope continues to trigger until memory is
filled, then the oscilloscope stops.
If the signal you are measuring has more than about 1 s of inactivity,
consider selecting the Normal trigger mode to prevent AutoTriggering. See
"To select the Auto or Normal trigger mode"
See Also
228
on page 229) to show infinite persistence across all acquired
"To set or clear persistence"
165 for details.)
segmented memory acquisitions.
number of segments into which you would like to divide the
oscilloscope's memory.
Memory can be divided into as few as two segments and as many as
1000 segments, depending on the oscilloscope model.
"Navigating Segments"
"Measurements, Statistics, and Infinite Persistence with Segmented
Memory"
on page 229
"Segmented Memory Re- Arm Time"

Chapter

on page 229
on page 230
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
on page 151 for details.
10, "Triggers," starting on page
on page 206.