UART/RS232 Triggering and Serial Decode
2 In the Search Menu, press the Search softkey; then, turn the Entry knob
3 In the Search Menu, press Search; then, select from these options:
For more information on searching data, see
page 146.
For more information on using the [Navigate] key and controls, see
"Navigating the Time Base"
470
to select the serial slot (Serial 1 or Serial 2) on which the UART/RS232
signals are being decoded.
Rx Data — Finds a data byte that you specify. For use when the DUT
data words are from 5 to 8 bits in length (no 9th (alert) bit).
Rx 1:Data — For use when the DUT data words are 9 bits in length
including the alert bit (the 9th bit). Finds only when the 9th (alert)
bit is 1. The specified data byte applies to the least significant 8 bits
(excludes the 9th (alert) bit)
Rx 0:Data — For use when the DUT data words are 9 bits in length
including the alert bit (the 9th bit). Finds only when the 9th (alert)
bit is 0. The specified data byte applies to the least significant 8 bits
(excludes the 9th (alert) bit).
Rx X:Data — For use when the DUT data words are 9 bits in length
including the alert bit (the 9th bit). Finds a data byte that you
specify regardless of the value of the 9th (alert) bit. The specified
data byte applies to the least significant 8 bits (excludes the 9th
(alert) bit).
Similar choices are available for Tx.
Rx or Tx Parity Error — Finds a parity error based on the parity you have
set in the Bus Configuration Menu.
Rx or Tx Any Error — Finds any error.
on page 75.
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
"Searching Lister Data"
on