Softkey
Source
Store Source to Arb
You can use the [Save/Recall] key and menu to save arbitrary waveforms to one of four
N O T E
internal storage locations or to a USB storage device, and you can recall them later. See
save arbitrary waveforms"

Creating New Arbitrary Waveforms

The New Waveform Menu is opened by pressing Create New in the Edit
Waveform Menu.
To create a new arbitrary waveform:
1 In the New Waveform Menu, press Initial Pts; then, use the Entry knob to
2 Use the Frequency/Frequency Fine/Period/Period Fine softkey to set the time
3 Use the Amplitude/High-Level and Offset/Low-Level softkeys to set the
4 When you are ready to create the new arbitrary waveform, press Apply
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
Description
Selects the analog channel or reference waveform to be captured and stored to the arbitrary
waveform. See
"Capturing Other Waveforms to the Arbitrary Waveform"
Captures the selected waveform source and copy it to the arbitrary waveform. See
"Capturing Other Waveforms to the Arbitrary Waveform"
select the initial number of points in the new waveform.
The new waveform will be a square wave with the number of points
you specify. The points are evenly spaced over the time period.
period bounding parameter (repetition frequency) of the arbitrary
waveform.
voltage bounding parameter of the arbitrary waveform.
& Edit.
on page 316 and
"To recall arbitrary waveforms"
Waveform Generator
on page 299.
on page 299.
on page 320.
17
"To
293