Calibration of new paper, emulsion changes (master paper)

1. Print the MBL on the previous (old) emulsion:
This provides for a calibrated condition of the
machine.
2. And right afterwards print a PBL without laser
point on the new master paper.
If a PBL with laser point is printed for the das MBL
paper, the previously valid MBL values will be
deleted, i.e. a PBL is also required for all other
papers.
If the old emulsion was used up before an MBL
could be printed, the calibration of the new paper
can only be started with a PBL. However, in this
case you should be sure that the machine is in a
calibrated condition. Otherwise there may be a
deviation between the master paper and all other
papers. If this is the case, a complete new
calibration is required.

Complete new calibration

If the papers are deviating too much a new
calibration is recommended:
1. Print an MBL on master paper
2. Directly afterwards print a PBL without laser
point for all papers other than the master
paper.
After manipulations on the laser module or after
changing the chemicals a new calibration is
recommended:
1. Print a PBL with laser point on the master
paper.
2. Print an MBL.
3. Directly afterwards print a PBL with laser point
for all papers other than the master paper.