Operation
Calibrating pumps
Call up:
–
Settings
Machine settings
PP pumps
Enter the water supply rate for DEV, BX, STB
–
Use a cylinder and hang in the hose elbow. The
–
pumping starts immediately when Activate xxx
pump is touched
Activate xxx pump
–
The pump button is only activated if a
calibration volume was entered.
When Readjust xxx is touched, a numeric
–
keypad for the entry of the measured volume is
opened. The pump is calibrated after entering a
value and pressing OK.
Reset to default is used to enter the default
–
values from the data base.
AGFA d-lab.3
2003-06-01 /PN 9008
Settings
015
4.29