New Feature: Compliance Film Calibration - AGFA DRYSTAR AXYS Service Manual

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
3.2.1

New Feature: Compliance Film Calibration

With software < 1.6.2_C1 the printer does not always pass the acceptance
procedure as defined by Afssaps / EUREF: The film calibration is not necessarily
accurate enough.
The "compliance film calibration" may be required as a last step of an acceptance
procedure that meets the special criteria set by Afssaps / EUREF.
It can be executed via new browser menu "Service Engineer Tools – Enter Service
command session – 6 Calibration – 6 Compliance Film Calibration".
For details refer to Technical Documentation Drystar Axys, DD+DIS033.07E,
Adjustments and Calibrations, edition 1, revision 1 (and higher revisions).
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
03-2011
Service - Calibration
1.
Film calibration
2.
Internal densitometer calibration
3.
Thermal print head calibration
4.
Horizontal alignment
5.
Smart profile calibration
6.
Compliance film calibration
Go back
This page is generated on 21-05-2008 12:33:33
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Software Releases and Patches
Service - Calibration
Compliance film calibration
Step Target values Measured values Step Target values Measured values
0.16
00
0.16
01
0.49
02
0.49
03
0.83
04
0.83
05
1.16
06
1.16
07
1.50
08
1.50
09
1.83
10
1.83
11
2.16
12
2.16
13
2.50
14
2.50
15
2.83
16
2.83
17
Start calibration
Go back
This page is generated on 22-05-2008 08:46:20
Figure 2
Agfa Company Confidential
0.33
0.33
0.66
0.66
0.99
0.99
1.33
1.33
1.66
1.66
2.00
2.00
2.33
2.33
2.66
2.66
3.00
3.00
Reset values
5367_SIB04_01.CDR
Chapter 3.8 / 10