Installation & Wiring
Wiring Procedures

Linear Voltage Inputs

2-12
SV100A/SV10AC User Guide

RTD Inputs

Use three wires having equal resistance to eliminate errors resulting from
lead length variations. Use 14 AWG wire with terminal connectors for long
lead runs and for 10-ohm copper RTDs. Refer to Figure 2–9 for
connection information and to the RTD manufacturer specifications to
determine color code connection information.
Figure 2-9.
Linear voltage inputs consist of variable voltage input ranges (±50, ±100,
±200 mV, ±1, ±5, ±10 V). Connect voltage inputs as shown in Figure 2–
10. Signal inputs greater than 10 V require the use of an input voltage
divider (consult your local representative or the factory).
Figure 2-10.