Wiring Procedures
Two-Wire Transmitter Power
2-14
SV100A/SV10AC User Guide
The optional two-wire transmitter power supply option provides an
isolated 24 Vdc current with a maximum output of 120 mA, which allows
Supply
you to power transmitters from the recorder rather than a separate power
supply. A maximum of six 4–20 mA or two 10–50 mA transmitters may be
powered from this source. Up to three 2-wire transmitter power supplies
are available, one on each input card. Refer to Figures 2–13 and 2–14 for
connection details. Two-wire transmitter inputs must have a 50-ohm shunt
resistor connected across the terminals.