Chart Speed; Trend Message; Display Previous - Thermo Scientific SmartView SV100A User Manual

100mm paperless data acquisition system
Table of Contents

Advertisement

The Function Menu

Chart Speed

Chart Speed
This menu item allows you to globally change the chart speed.
Figure 9-11.
Before you set a message to print on the trend screens and to the

Trend Message

event/alarm log, you must first program the messages, as detailed in Digital
I/O in Chapter 4. Once you have completed this step, return to this screen.
This function allows you to display the previous values for all point types

Display Previous

that can be reset. For example, you can display the value of a totalize point
type prior to the most recent point reset. Each previous value is shown
along with the point tag in individual boxes on the screen.
Figure 9-12.
Thermo Fisher Scientific
SV100A/SV10AC User Guide
9-5