Maintenance & Troubleshooting
General
Troubleshooting
Problem
Display totally black
Display is white or has
white lines
10-2
SV100/SV10C User Guide
Since the recorder is a complex instrument, we highly recommend that
troubleshooting and repair be performed only by a competent individual
(one capable of identifying existing and predictable hazards in the
surroundings or working conditions that are unsanitary, hazardous, or
dangerous to employees, and has authorization to make prompt corrective
measures to eliminate them).
Before beginning a detailed analysis of a malfunctioning recorder, we
recommend that you remove power from the recorder and perform a brief
visual inspection to determine if the problem is inclusive to the recorder.
Perform the following:
1. Ensure all cable connectors are seated properly.
2. Ensure all connections to the rear terminal panel are secure and wired
correctly (Chapter 2).
3. Use a VOM to ensure that the input voltage is applied to the operating
power terminal strip.
The following table lists some common problems and possible solutions.
Table 10-2. Troubleshooting
Checkpoint
Is power supply input voltage
correct?
Is power supply board
properly seated?
Does disk drive lamp
illuminate periodically?
Power supply board is
properly seated and disk drive
lamp does not illuminate.
Touch screen several times
and wait 10 seconds.
Yes/No
Solution
No
Connect recorder to proper input power
supply.
No
Reseat board.
Yes
Power supply board and fuse are OK.
Contact Thermo Scientific.
Check and/or replace fuse on power
supply board.
Replace power supply board, or
return recorder with power supply
module for repair.
If display returns to normal, configuration
file is corrupt. If display does not return to
normal, display module or CPU module
may be bad. Return for repair.