Thermo Scientific SmartView SV100A User Manual page 97

100mm paperless data acquisition system
Table of Contents

Advertisement

Thermo Fisher Scientific
Programming Parameters. Decimal Fix is also not programmable,
as log point are displayed as base 10 exponentials.
e. Press Next.
Figure 5–43. External type point programming: Display Scale screen
(log scale type)
2. Parameters screen:
Figure 5–44. External type point programming: Parameters screen
a. Input Scales block: Refer to step 4a in the Linear type point
programming section.
b. Output Scales block: Refer to step 4b in the Linear type point
programming section.
c. Timeout: Timeout sets a maximum time period between data
updates from the external source before flagging the point as
invalid. The point is flagged as invalid if no update is received
within the specified timeout period. Note that this invalid
condition may be assigned as an abnormal alarm to annunciate loss
of communications with the Host. Enter the timeout period in
seconds or enter 0 to disable this function.
d. Press Next.
Point Programming
Programming
SV100A/SV10AC User Guide
5-39