Restoring Calibration Constant Defaults - Thermo Scientific SmartView SV100A User Manual

100mm paperless data acquisition system
Table of Contents

Advertisement

Calibration

Restoring Calibration Constant Defaults

Restoring
Calibration
Constant Defaults
8-4
SV100A/SV10AC User Guide
Note The mA reading may not show the correct value until current range
calibration is completed and you return to the calibration screen. ▲
Three constants are calculated and saved in nonvolatile memory for each
voltage range, and one constant is calculated and saved in nonvolatile
memory for the current source. ADC measurement operation can be
restored to the factory defaults if the calibration constants have been altered
or lost. If the constants in memory are lost and/or the ADC measurement
seems inaccurate, follow the procedure below to restore factory defaults.
Note This procedure is not a calibration procedure for your instrument. It
is a procedure to restore the ADC constants from the original factory
calibration. Do not perform this procedure if calibration equipment is
available to calibrate your instrument. ▲
1. From the Program menu, select Measurement > ADC Constants.
2. Select the ADC board for which you want to restore constant defaults.
3. Press Recall. Constant defaults are reloaded for that board.