The Program Menu
Media Control
4-6
SV100A/SV10AC User Guide
for the particular points' stored data. The formula for determining
the boundary setting follows:
Boundary = Output Data Span  Desired Accuracy.
Example: Span of 0 to 20; accuracy of 0.1%
Boundary = 20  0.001 = 0.02.
The boundary can be set at 0.02, and any signal variation outside
this range from the previous signal is stored to disk. Any signal not
going outside of this range is ignored and not stored. Compression
is dependent upon the fluctuation of the measured signal. A large
amount of signal fluctuation results in less compression and a larger
disk space requirement. Small signal fluctuation results in more
compression and a smaller disk space requirement. The accuracy of
the stored signal readings is maintained at the level you set.
Note Files must be Encrypted when using the Pronto Software Version
5.71 or newer. ▲
b. ASCII: ASCII storage mode allows you to open the data files in a
spreadsheet or a word processing program. However, ASCII format
requires about twice the media space as compressed storage.
3. Media Mode:
a. Fill: Storage media fills to the end with data.
Figure 4–7.
Table 4-3.
Fill Option
Fill Contact
Fill Percent
Description
Assign the standard alarm contact used to signal a disk full
alarm.
Set a percentage of disk space the recorder fills before
activating a low disk space alarm.