2. Display Scale screen:
Figure 5–30. Gated Timer Calculated type point programming: Display Scale
screen
a. Text and Colors blocks: Refer to the Similar Programming
Parameters section for instructions.
b. Display block: Refer to the Similar Programming Parameters
section for instructions. Note that the only available Decimal Fix
selection is 0.
c. Press Next.
3. Parameters screen:
Figure 5–31. Gated Timer Calculated type point programming: Parameters
screen
a. Gate: If a base point of 0 is specified, the gated timer runs
continuously until reset. If a valid point (1–36) is assigned as the
base point, the timer accumulates seconds only when the base point
evaluates to be true.
b. Reset Control: Refer to Reset Control in the High Peak, Low Peak,
Timed Average section for instructions.
c. Press Next.
Point Programming
Programming
SV100A/SV10AC User Guide
5-29