Thermo Scientific SmartView SV100A User Manual page 88

100mm paperless data acquisition system
Table of Contents

Advertisement

Point Programming
Programming
5-30
SV100A/SV10AC User Guide
4. Alarm Limits screen: Refer to the Similar Programming Parameters
section for instructions.
5. Press Apply
 Sterilize: The sterilize type point allows you to set a reference
temperature and a Z constant to calculate the value of a point. The
calculation used follows:
F
= F
0
0-1
where:
BP = Base point
Ref = Reference temperature
Z = Z constant (default value is 10.0°C)
1/60 = Calculation is performed once per second
The calculation can be used for both F
reference temperature and the Z constant.
Note Sterilize type points are not supported in the current release. Please
contact the factory. ▲
1. Type screen: Select Sterilize > Next.
Figure 5–32. Sterilize Calculated type point programming: Type screen
2. Display Scale screen: Refer to the Similar Programming Parameters
section for instructions.
3. Parameters screen:
+ (1/60) * 10 ^ ((BP – Ref) / Z),
and F
0
by changing the
n