Calibration
Scale Calibration
Procedure
8-2
SV100A/SV10AC User Guide
1. Connect the precision voltage source to the positive (+) and negative (-)
calibration inputs. Verify correct polarity.
Figure 8-1. Connection detail
2. Turn on the voltage source and allow it to stabilize.
3. Press Program > Measurement.
4. Press ADC Calibration to access the calibration screen.
Figure 8-2.
5. Press the Input Number text box and enter the input number that has
the precision voltmeter connection. Press OK.
6. Select the range to calibrate (50 mV, 100 mV, 200 mV, 1 V, 5 V, 10
V).