Calculations are performed on the instantaneous value of one or more

Calculated

points. The calculations are performed after each scan of all measured
inputs.
 High Peak, Low Peak, Time Average: High peak calculated type points
track the highest data of a given point. This data is stored until some
form of reset occurs. When the data is reset, the high peak point data is
printed on the chart, along with the time the peak value occurred. Low
peak calculated type points track the lowest data of a given point. This
data is stored until some form of reset occurs. When the data is reset,
the low peak point data is printed on the chart, along with the time the
peak value occurred. Time average calculated type points calculate a
continuous average of the measured or processed value of a given point.
The value of the average at the time of reset represents the true average
of the base points' data since the previous reset time.
1. Type screen: Select High Peak, Low Peak, or Time Average. Press
Next to continue.
Figure 5–21. Hi Peak Calculated type point programming: Type screen
2. Display Scale screen: Refer to the Similar Programming Parameters
section for instructions.
3. Parameters screen:
Point Programming
Programming
SV100A/SV10AC User Guide
5-19