Sorvall ST 8 Centrifuge
Instruction Manual
50138912-a
February 2013

Advertisement

Table of Contents
loading

Summary of Contents for Thermo Scientific Sorvall ST 8

  • Page 1 Thermo Scientific Sorvall ST 8 Centrifuge Instruction Manual 50138912-a February 2013 Visit us online to register your warranty. www.thermoscientific.com/labwarranty...
  • Page 3 Thermo Scientific Sorvall ST 8 Centrifuge Instruction Manual 50138912-a February 2013...
  • Page 4 © 2013 Thermo Fisher Scientific Inc. All rights reserved. Delrin, TEFLON, and Viton are registered trademarks of DuPont. Noryl is a registered trademark of SABIC. POLYCLEAR is a registered trademark of Hongye CO., Ltd. Hypaque is a registered trademark of Amersham Health As.
  • Page 5 WEEE Compliance This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is marked with the following symbol: Thermo Fisher Scientific has contracted with one or more recycling or disposal companies in each European Union (EU) Member State, and these companies should dispose of or recycle this product.
  • Page 6: Table Of Contents

    Description ........... . 41 Thermo Scientific Sorvall ST 8 Centrifuge...
  • Page 7 ..............65 Sorvall ST 8 Centrifuge...
  • Page 8: Preface

    Items Supplied Item Quantity Check Thermo Scientific Sorvall ST 8 Centrifuge Power supply cable Instruction manual If any parts are missing, please contact your nearest Thermo Fisher Scientific representative. Intended Use This centrifuge is a laboratory product used to separate substance mixtures of different densities.
  • Page 9 This symbol refers to general hazards. CAUTION means that material damage could occur. WARNING means that injuries or material damage or contamination could occur. This symbol refers to biological hazards. This symbol refers to electrical hazards. Sorvall ST 8 Centrifuge Thermo Scientific...
  • Page 10: Precautions

    • If a hazardous situation occurs, turn off the power supply to the centrifuge and leave the area immediately. Note In order to ensure safe operation of the Thermo Scientific Sorvall ST 8 Centrifuge, general safety regulations must be followed: Mind the regulations in your country.
  • Page 11 Preface WARNING In order to ensure safe operation of the Thermo Scientific Sorvall ST 8 Centrifuge, the following general safety regulations must be followed: • The centrifuge should be operated by trained specialists only. • The centrifuge is to be used for its intended use only.
  • Page 12 • Highly corrosive substances which can cause material damage and impair the mechanical stability of the rotor, should only be centrifuged in corresponding protective tubes. Note If used for Foodstuffs machinery, for cosmetics or pharmaceutical products, only use closed or aerosol-tight containers for centrifugation. Thermo Scientific Sorvall ST 8 Centrifuge...
  • Page 13 Preface Sorvall ST 8 Centrifuge Thermo Scientific...
  • Page 14: Introduction And Description

    Introduction and Description Contents “Characteristics” page 16 • “Technical Data” page 17 • “Directives, Standards and Guidelines” page 18 • “Functions and Features” page 19 • “Mains Supply” page 19 • “Rotor Selection” page 20 • Thermo Scientific Sorvall ST 8 Centrifuge...
  • Page 15: Characteristics

    Introduction and Description Characteristics Characteristics The Thermo Scientific Sorvall ST 8 Centrifuge is an in-vitro diagnostics device according to the In-Vitro Diagnostics Directive 98/79/EC. Several rotors with a wide range of tubes can be used. The set speed is reached within seconds. The maintenance-free induction motor ensures quiet and low-vibration operation even at high speeds, as well as guaranteeing a very long lifetime.
  • Page 16: Technical Data

    Introduction and Description Technical Data Technical Data The technical data of the Thermo Scientific Sorvall ST 8 Centrifuge is listed in the following table. Tabelle 1-1. Technical Data Thermo Scientific Sorvall ST 8 Centrifuge Feature Value Environmental Conditions Designed and intended for indoor use Altitudes of up to 2,000 m above Sea Level max.
  • Page 17: Directives, Standards And Guidelines

    IEC 61010-2-101, 2nd Edition EN ISO 13485 Japan IEC 61010-1, 2 Edition 100V 60Hz IEC 61010-2-020, 2 Edition 100V 50Hz IEC 61010-2-101, 2 Edition China EN 61326-1 Class B 230V 50/60Hz EN 61326-2-6 EN ISO 13485 Sorvall ST 8 Centrifuge Thermo Scientific...
  • Page 18: Functions And Features

    Mains Supply The following table contains an overview of the electrical connection data for the Thermo Scientific Sorvall ST 8 Centrifuge. This data is to be taken into consideration when selecting the mains connection socket. Tabelle 1-4. Electrical connection data of the Thermo Scientific Sorvall ST 8 Centrifuge...
  • Page 19: Rotor Selection

    Introduction and Description Rotor Selection Rotor Selection The Thermo Scientific Sorvall ST 8 Centrifuge is supplied without a rotor. Various rotors are available to choose from. Thermo Scientific Rotors Part No. TX-150 swinging bucket rotor 75005701 TX-150 round buckets 75005702...
  • Page 20: Before Use

    Before Use Contents “Before Setting up” page 22 • “Transporting the Centrifuge” page 23 • “Location” page 22 • “Aligning the Centrifuge” page 23 • “Mains Connection” page 24 • “Storage” page 24 • Thermo Scientific Sorvall ST 8 Centrifuge...
  • Page 21: Before Setting Up

    WARNING UV rays reduce the stability of plastics. Do not subject the centrifuge, rotors and plastic accessories to UV rays like in direct sunlight. • The set-up location must be well-ventilated at all times. Sorvall ST 8 Centrifuge Thermo Scientific...
  • Page 22: Transporting The Centrifuge

    The supporting structure must be suitable for horizontal setup of the centrifuge. CAUTION If the centrifuge isn’t leveled, imbalances can occur and the centrifuge can be damaged. Do not place anything under the centrifuge feet to level the centrifuge. Thermo Scientific Sorvall ST 8 Centrifuge...
  • Page 23: Mains Connection

    • Before storing the centrifuge and the accessories, it must be cleanend, and if necessary, desinfected and decontaminated. • Store the centrifuge in a clean, dust-free location. • Be sure to place the centrifuge on their centrifuge feet. • Avoid direct sunlight. Sorvall ST 8 Centrifuge Thermo Scientific...
  • Page 24: Shipping The Centrifuge

    • The decontamination must be confirmed in a special form. WARNING Before shipping the centrifuge and accessories you have to clean and if necessary disinfect or decontaminate the entire system. In doubt contact the Thermo Fisher Scientific customer service. Thermo Scientific Sorvall ST 8 Centrifuge...
  • Page 25 Before Use Shipping the Centrifuge Sorvall ST 8 Centrifuge Thermo Scientific...
  • Page 26: Control Panel

    Control Panel Contents “Control Panel” page 28 • “Keys” page 28 • Thermo Scientific Sorvall ST 8 Centrifuge...
  • Page 27: Control Panel

    Use the ARROW keys in order to modify the displayed value Change Value Press the ACCELERATION/DECELARATION key multiple times to change the Acceleration/ different profiles Decelaration Use the TOGGLE key to change the display mode. (Speed / RCF Value) Changing the display mode Sorvall ST 8 Centrifuge Thermo Scientific...
  • Page 28: Operation

    31 • “Entering Parameters” page 32 • “Saving Programs” page 34 • “Centrifugation” page 35 • “Short-term Centrifugation” page 37 • “Removing the Rotor” page 37 • “Aligning the Centrifuge” page 38 • Thermo Scientific Sorvall ST 8 Centrifuge...
  • Page 29: Switch On Centrifuge

    “Mechanical Emergency Door Release” page 54). Close Lid Close the lid by pressing down on it lightly in the middle or on both sides of it. Note The lid should audibly click into place. Sorvall ST 8 Centrifuge Thermo Scientific...
  • Page 30: Rotor Installation

    Operation Rotor Installation Rotor Installation The approved rotors for the Thermo Scientific Sorvall ST 8 are listed in section “Rotor Selection” page 20. Use only the rotors and accessories from this list in the centrifuge. CAUTION Unapproved or incorrectly combined accessories can cause serious damage to the centrifuge.
  • Page 31: Entering Parameters

    4. Close the centrifuge lid. Entering Parameters The Thermo Scientific Sorvall ST 8 offers you 2 profiles: standard and soft. Press the ACCELERATION / DECELERATION key to set a profile. The LEDs show the chosen settings. The last profile is saved, if you restart the centrifuge.
  • Page 32 The maximum RCF value is related to the maximum radius of the tube opening. Remember that this value is reduced depending on the tubes and adapters used. This can be accounted for in the calculation above if required. Thermo Scientific Sorvall ST 8 Centrifuge...
  • Page 33: Programs

    During continuous operation, the centrifuge will continue running until you stop it manually. Programs The Thermo Scientific SL 8 Centrifuge is able to save up to 4 programs. It is only possible to save a program, if the centrifuge is not running.
  • Page 34: Centrifugation

    You cannot open the lid as long as the centrifuge is running. Imbalance Indicator If a load is imbalanced, this will be indicated at speed higher than approx. 300 rpm by the message "Imbalanced load". The run will terminate. Thermo Scientific Sorvall ST 8 Centrifuge...
  • Page 35 Press the STOP key on the control panel. The centrifuge will be decelerated at the designated rate. The message RUN COMPLETED will illuminate, and after pressing the OPEN key, the lid will open and you can remove the centrifuged samples. Sorvall ST 8 Centrifuge Thermo Scientific...
  • Page 36: Short-Term Centrifugation

    Short-term Centrifugation Short-term Centrifugation For short-term centrifugation, the Thermo Scientific Sorvall ST 8 has a PULSE-function. By holding down the PULSE key, spinning will start and continue until the key is let go. The centrifuge accelerates and brakes at maximum power. Any rpm or RCF entered beforehand is overridden.
  • Page 37: Aerosol-Tight Rotors

    Figure 4. Auto-Lock-lid for aerosol-tight rotors WARNING Mind the Auto-Lock-mandrel inside the lid. Do not touch the mandrel. Aligning the Centrifuge • To turn off the centrifuge push the mains switch to "0". Sorvall ST 8 Centrifuge Thermo Scientific...
  • Page 38: System Menu

    To get into the system menu, press and hold any button when turning on the centrifuge until the system menu shows up in the display. Use the ARROW keys below the time selection in order to navigate within the system menu points. Thermo Scientific Sorvall ST 8 Centrifuge...
  • Page 39 System Menu English Nederlands Росский Francais Español Italiano Deutsch 0602 V01 4758 V05 Software and NVRAM identification exam- ple values. Sorvall ST 8 Centrifuge Thermo Scientific...
  • Page 40: Description

    Use the ARROW keys below SPEED in order to navigate through the system menu. Press the START key to save this edit and quit the system menu. Press the STOP key to quit the system menu. Thermo Scientific Sorvall ST 8 Centrifuge...
  • Page 41 Use the ARROW keys below the speed display in order to navigate through the system menu. Press the START key to save this edit and quit the system menu. Press the STOP key to quit the system menu. Sorvall ST 8 Centrifuge Thermo Scientific...
  • Page 42 Use the ARROW keys below SPEED in order to navigate through the system menu. Press the START key to save this edit and quit the system menu. Press the STOP key to quit the system menu. Thermo Scientific Sorvall ST 8 Centrifuge...
  • Page 43 System Menu Description Sorvall ST 8 Centrifuge Thermo Scientific...
  • Page 44: Maintenance And Care

    Maintenance and Care Contents “Cleaning Intervals” page 46 • “Cleaning” page 46 • “Disinfection” page 48 • “Decontamination” page 49 • “Autoclaving” page 50 • “Thermo Fisher Scientific Service” page 50 • Thermo Scientific Sorvall ST 8 Centrifuge...
  • Page 45: Cleaning Intervals

    • After cleaning, treat the entire surface of aluminum parts with corrosion protection oil (part no. 70009824). Also treat the cavities with oil. • Store the aluminum parts at room temperature or in a cold-storage room with the cavities pointing down. Sorvall ST 8 Centrifuge Thermo Scientific...
  • Page 46 CAUTION When cleaning, do not allow liquids, especially organic solvents, to get on the drive shaft or the bearings of the centrifuge. Organic solvents break down the grease in the motor bearing. The drive shaft could lock Thermo Scientific Sorvall ST 8 Centrifuge...
  • Page 47: Disinfection

    • After cleaning, treat the entire surface of aluminum parts with corrosion protection oil (part no. 70009824). Also treat the cavities with oil. • Tread the bolt of the swing out rotor with bolt grease (part no. 75003786). Sorvall ST 8 Centrifuge Thermo Scientific...
  • Page 48: Decontamination

    • After cleaning, treat the entire surface of aluminum parts with corrosion protection oil (part no. 7000 9824). Also treat the cavities with oil. • Tread the bolt of the swing out rotor with bolt grease (part no. 75003786). Thermo Scientific Sorvall ST 8 Centrifuge...
  • Page 49: Autoclaving

    Thermo Fisher Scientific offers inspection and service contracts for this work. Any necessary repairs are performed for free during the warranty period and afterwards for a charge. This is only valid if the centrifuge has only been mainained by a Thermo Fisher Scientific service technician. Sorvall ST 8 Centrifuge Thermo Scientific...
  • Page 50: Disposal

    For the countries of the European Union the disposal is regulated by the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. See “WEEE Compliance” page 5 Mind the information on transport and shipping. See “Transporting the Centrifuge” page 23 “Shipping the Centrifuge” page 25 for more information. Thermo Scientific Sorvall ST 8 Centrifuge...
  • Page 51 Disposal Sorvall ST 8 Centrifuge Thermo Scientific...
  • Page 52: Troubleshooting

    Troubleshooting Contents “Mechanical Emergency Door Release” page 54 • “Troubleshooting by Guide” page 55 • “When to Contact Customer Service” page 57 • Thermo Scientific Sorvall ST 8 Centrifuge...
  • Page 53: Mechanical Emergency Door Release

    4. Push the cord back into the centrifuge and mount the plug. Reconnect the centrifuge once the power has been restored. Switch on the centrifuge. Press the OPEN key to have the door locks operative again. Sorvall ST 8 Centrifuge Thermo Scientific...
  • Page 54: Troubleshooting By Guide

    Let the centrifuge cool down for 15 minutes. If the error message appears again, inform the customer service. E-017; Read Manual Wait until the rotor has stopped. E-020; Check, if the rotor is qualified for the Sorvall ST 8 Centrifuge (check “Rotor Selection” page 20). E-021;...
  • Page 55 If the error message appears again, inform the customer service. E-029; Read Manual Check, if a rotor is installed. E-045 Check, if the rotor is qualified for the Sorvall ST 8 Centrifuge (check “Rotor Selection” page 20). Restart the centrifuge.
  • Page 56: When To Contact Customer Service

    For a desription how to get there, see “Enter system menu” page 41. You will find a description on how to find the “Software ID” page 43 and for the “NVRAM ID” page Thermo Scientific Sorvall ST 8 Centrifuge...
  • Page 57 Troubleshooting When to Contact Customer Service Sorvall ST 8 Centrifuge Thermo Scientific...
  • Page 58: Chemical Compatibility Chart

    S S U U S U M U S U U S U U U M U M U U U S U U S U S Benzyl Alcohol U U - M M - S U U U U U U U - M S M - Thermo Scientific Sorvall ST 8 Centrifuge...
  • Page 59 S U S S S S S S S S M S M S Ethylene Glycol Ethylene Oxide Vapor S U - S M - S S S U S U S S S U Sorvall ST 8 Centrifuge Thermo Scientific...
  • Page 60 S U U S M S U S S S S Mercaptoacetic Acid Methyl Alcohol S S S U S S M S S S S S U S U M S S S S S S S M S M U Thermo Scientific Sorvall ST 8 Centrifuge...
  • Page 61 M S U U M M S S S M S M S U M U - Phosphoric Acid (conc.) Physiologic Media (Serum, Urine) M S S S - S S S S S S S S S S S S S S S S S S S Sorvall ST 8 Centrifuge Thermo Scientific...
  • Page 62 S S U S S S M U S U U U U U U M U M U U U S U M S U S U U S S S S U S S S S S S S S S S S S S S S S U S S S Zinc Chloride Thermo Scientific Sorvall ST 8 Centrifuge...
  • Page 63 No data available. Because no organized chemical resistance data exists for materials under the stress of centrifugation, when in doubt we recommend pretesting sample lots. suggest testing, using sample to avoid loss of valuable material. Chemical resistance data is included only as a guide to product use. Sorvall ST 8 Centrifuge Thermo Scientific...
  • Page 64: Contact

    +91 22 6716 2200 Japan +81 45 453 9220 Other Asian Countries +852 2885 4613 Australia +61 39757 4300 New Zealand +64 9 980 6700 Countries not listed +49 6184 90 6000 +33 2 2803 2180 Thermo Scientific Sorvall ST 8 Centrifuge...
  • Page 65 Contact Sorvall ST 8 Centrifuge Thermo Scientific...
  • Page 66 Failure message 55 Technical Data 17 Features 19 Thermo Fisher Scientific Service 50 Transporting the Centrifuge 23 Trouble Shooting 53 Imbalance indicator 35 WEEE Compliance 5 Keys 28 Kundendienst 50, 57 Lid Opening 30 Location 22 Thermo Scientific Sorvall ST 8 Centrifuge...