Point Programming

Programming

Programming
5-2
SV100A/SV10AC User Guide
 Scaling: The outputs of the conditioning block are the values in the
digital windows and are represented in common engineering units. The
outputs of the conditioning block are also sent to a scale block where a
chart scale is applied to each point. The display scales determine what
part of the full scale range is used by the display/record block.
 Display/record: These scale points are then applied to the
display/record block that consists of the visual information or recorded
information that the user requires. A maximum of six scaled points may
be assigned to each bargraph and trend display type. All may be
recorded to disk.
Note that applying scaling other than full scale output to recorded data
limits the data to the range between scale end points. It is possible to
store the base point with different scaling than is used to display it on
the chart by using one of the computational points.
The recorder can be programmed remotely; however, the following sections
apply to manually programming the unit from the front panel using the
display and the touch screen.
Once you have configured a point, save it by pressing the Apply button on
that screen. When you exit a programming session, press Apply on the
main Point Programming screen to save the new parameters in nonvolatile
memory.
Figure 5–1.
1. Touch the Start Point number box, and a Point Set input box appears.
Enter the point number you wish to program or the starting number
for a range of points.
Continue to step 2 if you are programming a range of points.
Otherwise, continue to the next section Similar Programming
Parameters.