Point Programming
Programming
5-34
SV100A/SV10AC User Guide
Figure 55–36. Equation Calculated type point programming: Display Scale
screen (log scale)
3. Parameters screen:
Figure 5–37. Equation Calculated type point programming: Parameters screen
Create or modify an equation by pressing the Edit button and using
the equation editor that appears (Figure 5–38). Press Accept when
complete.
Figure 5–38. Equation entry keypad
A total of 40 operators and operands may be used in an equation. If an
equation is not constructed properly, an Error: Invalid Data dialog box
displaying Illegal Equation appears.