Thermo Scientific SmartView SV100A User Manual page 90

100mm paperless data acquisition system
Table of Contents

Advertisement

Point Programming
Programming
5-32
SV100A/SV10AC User Guide
point is reset and begins again. If the sterilization point is
bypassed and reactivated, the point resets and begins again.
iv. Target Temperature: Enter the minimum allowable
temperature for the process.
v. Deadband: This is the allowable range above and below the
target temperature.
d. Reference Temperature: The reference temperature is the mean
temperature that the process must achieve and hold.
e. Press Next.
4. Alarm Limits screen: Refer to the Similar Programming Parameters
section for instructions.
5. Press Apply
 Equation: This type of point allows you to program an equation to
calculate a value of one or more points and/or constants.
1. Type screen: Press Equation > Next.
Figure 5–34. Equation Calculated type point programming: Type screen
Thermo Fisher Scientific

Advertisement

Table of Contents
loading

This manual is also suitable for:

Smartview sv100ac

Table of Contents