automatically measured on the 1 V range) and 10–50 mA (assumes the
use of a 100-ohm shunt resistor and is automatically measured on the 5
V range).
If (a) current inputs are not predefined, (b) it is desired to use any
selectable voltage range, and/or (c) it is desired to use a different shunt
resistor for a current input, use Ohm's Law to determine the
appropriate low and high end input voltages:
E = IR
where:
E = Equivalent voltage inputs
I = Current in mA
R = Precision shunt resistance value
The calculation can be used for both F
Figure 5–9. Linear type point programming: Type screen
2. Display Scale screen: Refer to the Similar Programming Parameters
section for instructions. When complete, press Next.
Point Programming
and F
by changing the
0
n
SV100A/SV10AC User Guide
Programming
5-9