Chapter 13 - Specifications
O
, CO
, and Anesthetic Gas Measure-
2
2
ment
Side-stream sampling
(the sampled flow is returned to
the breathing system and taken into account for
measurement and delivery)
calibration conditions ATPS, sampling rate in NTPD.
Sampling rate
Delay for sampling (typical value, depends on
sample line)
Response time t
..
O
10
90
2
Response time t
..
CO
10
90
Response time t
..
anesthetic agents
10
90
O
Measurement
2
(consumption-free, paramagnetic measurement)
Measuring range
Resolution of the measurement
Resolution of the indication
(for ins. O
, exp. O
)
2
2
Accuracy
CO
Measurement
2
(infrared spectrometry)
Measuring range
Resolution of the measurement
Resolution of the indication
(for inCO
, etCO
)
2
2
Accuracy
Anesthetic Gas Measurement
(infrared spectrometry)
Measuring range, anesthetic agent
Halothane
Isoflurane
Enflurane
Sevoflurane
Desflurane
Resolution of the measurement
Resolution of the indication
(for insp. and exp. anesthetic agent)
Accuracy (at respiration rates of up to
60 bpm and I:E ratio of 1:1)
216
. All values measured under
200 mL/min ±20 mL/min
less than 4 seconds
(ambient temp.>68°F)
less than 500 ms
less than 350 ms
2
less than 500 ms
0 to 100 Vol.%
0.1 Vol.%
1 Vol.%
±3 Vol.% in the measuring range 0 to 100 Vol.%
0 to 76 mmHg
1 mmHg
1 mmHg
±3.8 mmHg or ±12% of measured value, whichever is greater
All values in Vol.% refer to ambient pressure 760 mmHg
0 to 8.5 Vol.%
0 to 8.5 Vol.%
0 to 10 Vol.%
0 to 10 Vol.%
0 to 20 Vol.%
0.1 Vol.%
0.1 Vol.%
±(0.15 Vol.% + 15% rel.)
Specifications