What To Do If Problems Arise; Additional Maintenance And Care Instructions - Dräger Cato Instructions For Use Manual

Anaesthetic workstation
Table of Contents

Advertisement

Descriptions
Disconnecting the Cato ceiling version

What to do if problems arise

Additional maintenance and care instructions

1 Move Cato away from the DVE.
2 Swing supporting flap upwards.
The DVE can now be controlled via its own keypad
again!
As from approx. 5 cm above the lower limit position,
the DVE can also be moved upwards even when the
supporting flap is lowered.
What to do if problems arise
If the machine inadvertently strikes an obstacle before reaching the floor, the DVE will stop before it is forcibly
disconnected and cannot be controlled via the keypad on the Cato.
The machine can then only be raised via the keypad on the DVE.
If the machine subsequently remains at an angle on its mounting journal:
Briefly lower it onto the ground and pick it up again with the DVE.
The mount may suffer damage in extreme cases!
Call DrägerService if damage is suspected. Depending on how serious the damage is, the Cato may crash to
the ground when raised again!
Additional maintenance and care instructions
The windows on the optical sensors must be cleaned!
Abrasives and scratching agents must not be used.
The sensors are resistant to all cleaning agents and disinfectants usually used in hospitals.