Carry out machine cleaning and disinfection
1
Observe the instructions for use for the washer-
disinfector.
2
Position items so that all interior spaces are com-
pletely flushed and water can drain off freely.
3
Use a suitable cleaning agent.
4
Select suitable program (preferably anesthesia
program).
Cleaning must be carried out at 40 °C to
60 °C (104 °F to 140 °F) for at least 5 min-
utes.
Thermal disinfection must be carried out at
80 °C to 95 °C (176 °F to 203 °F) and with
corresponding contact time.
5
Carry out final rinsing with deionized water.
6
Immediately remove items from the washer-dis-
infector.
7
Check parts for visible dirt and damage. If neces-
sary, repeat program or carry out manual clean-
ing and disinfection.
8
Allow items to dry thoroughly.

Sterilization

During sterilization, living microorganisms are re-
moved from semi-critical medical devices. Residual
water inside the components is also dried out.
 Only sterilize cleaned and disinfected items.
Use a vacuum steam sterilizer (in accordance with
EN 285 / EN 13060), preferably with fractional vacu-
um.
Instructions for use Inhalette

Visual inspection

 Inspect all items for damage and wear, e.g.
cracking, embrittlement or pronounced harden-
ing, and residual soiling.
CAUTION
Even accessories designed to be reused have a
limited service life. Handling and reprocessing can
increase wear and markedly shorten service life
(e.g., disinfectant residues can attack the material
more intensely during autoclaving). If signs of wear
become visible, such as cracks, deformation, dis-
coloration, peeling, etc., affected accessories must
be replaced.

Service

This chapter describes the maintenance measures
required to maintain the proper functioning of the
medical device. Maintenance measures must be
performed by the personnel responsible.
WARNING
Risk of infection
Users and service personnel can become in-
fected with pathogenic germs.
Disinfect and clean device or device parts be-
fore any maintenance measures and also be-
fore returning the medical device for repair.

Definition of service terminology

Con-
Definition
cept
Service
All measures (inspection, repair) intend-
ed to maintain and restore the functional
condition of a medical device
Inspec-
Measures intended to determine and
tion
assess the actual state of a medical de-
vice

Repair

Measures intended to restore the func-
tional condition of a medical device after
a device malfunction

Inspection

Perform inspections at regular intervals and observe
the following specifications.
.
Checks
Interval
Personnel re-
sponsible
Inspection
Every 12
Service personnel
months
Perform inspection
1
Check accompanying documents:
Instructions for use are available
2
Check correct functioning in accordance with
"Checking correct functioning" on page 15.
Repair
Dräger recommends that all repairs are performed
by DrägerService and that only authentic Dräger re-
pair parts are used.