Application

Intended use

WARNING
Risk of device failure and/ or danger to patient
If the intended use of this anesthesia machine
is not adhered to, it may fail and/or the patient
may be endangered.
Use the anesthesia machine only as specified
in the intended use of these instructions for
use.
Primus Infinity Empowered – Anesthesia worksta-
tion for adults, children, and neonates with auto-
matic and manual ventilation, as well as spontane-
ous breathing with or without pressure support.

Utilization

– Inhalation anesthesia in rebreathing systems
– Inhalation anesthesia in semi-closed to virtually
closed systems with "low flow" and "minimal
flow" techniques (for minimal gas and anes-
thetic agent consumption)
– Inhalation anesthesia in non-rebreathing sys-
tems, with a separate fresh-gas outlet for con-
necting, e.g., a Bain system or Magill system,
with a fresh-gas flow of 0.2 to 18 L/min (op-
tional).
Optional:
– Operation without nitrous oxide: The device
may only be operated with O

Ventilation modes

– Volume-controlled ventilation in Volume Mode.
Optional activation of: Synchronization,
Press. Support (Pressure support) (optional)
– Pressure-controlled ventilation in Pressure
Mode
16
– Manual ventilation (Man.)
– Spontaneous breathing (Spont.)
– Pressure-supported spontaneous breathing
– Volume AF (Volume Mode AutoFlow) (optional)
The following measured values are dis-
played
– Peak pressure PEAK, mean pressure P
– Expiratory minute volume MV,
– patient compliance C
– Inspiratory and expiratory concentration of O
and air.
2
Optional:
– Functional oxygen saturation (SpO
Instructions for use Primus Infinity Empowered SW 4.5n
Optional activation of: Synchronization,
Press. Support (Pressure support) (optional)
Press. Support
CPAP
Optional activation of: Synchronization,
Press. Support (Pressure support) (optional)
plateau pressure PLAT, and PEEP
difference between insp. and exp. minute vol-
ume MV
,
LEAK
PAT
tidal volume V
,
T
breathing rate freq.
N
O, anesthetic gas, and CO
2
O
:
2
Difference between inspiratory and expiratory
O
concentration
2
rate
(optional)
MEAN,
,
2
) and pulse
2
,
2

Advertisement