Advertisement

CO
measurement (optional)
2
The CO
measurement only works if the CO
2
option has been installed and if the CO
present.
1 Connect the CO
sensor and the cuvette (refer
2
to "Connecting the CO
on page 48).
2 Set the cuvette type in the Settings menu (refer
to Cuvette type setting).
The following will be activated:
– Curve window: CO
– Measured values window: the parameter etCO
will automatically be displayed
– Alarm window: etCO
For more information on curves and measured
values: refer to "CO
curve large view" on page 34.
2
For more information on configuring measured
value pairs: refer to "Customer Service Mode"
on page 98.
For connecting the CO
page 48. For CO
zero calibration and filter check
2
before ventilation, see page 58. For CO
tion in the Customer Service Mode see page 105.
Instructions for Use Oxylog 3000 plus SW 1.n
2
sensor is
2
sensor and the cuvette"
2
Curve
2
high and etCO
low alarm
2
2
sensor and cuvette see
2
configura-
2

Cuvette type setting

NOTE
The cuvette windows of the reusable cuvette and
disposable cuvette have different optical proper-
ties. Therefore, the correct cuvette type must be
selected in the Setting menu. Otherwise the zero
point is shifted by up to ± 8 mmHg of CO
To set the cuvette type (reusable or disposable):
1 Press the Settings
2 Select and activate the line Cuvette type.
3 Set the cuvette type and confirm.
2
NOTE
If a wrong cuvette type is selected, the Oxylog
3000 plus displays the alarm !!! Check cuvette
type.
Checking the CO2 sensor during
ventilation
The following checks are necessary for the CO
sensor during ventilation:
Check
Interval
CO
zero cali-
Recommended before measure-
2
bration
ment and when changing the
CO
CO
filter
Required in intervals of one
2
check
month.
Monitoring
.
2
key.
sensor to another unit.
2
2