Page 1
The Oxylog 2000 plus Pocket Guide is not a replacement ® Oxylog 2000 plus Pocket Guide or substitute for the Instructions for Use. Any use of the device requires full understanding and strict observation Software version 1.n of the Instructions for Use. The user must be familiar with the device according to the national and local laws and recommendations.
Index Front panel with all options ........6-7 Side view, right / Rear view .
Page 4
Index Determining the approximate pneumatic operating time for the Oxylog® 2000 plus .....22-23 Operation / Settings Starting operation ......... . .24 Shutdown .
Page 5
Index Alarms and Measured Values In the event of an alarm ........38 Setting alarm limits .
Front panel with all options The Oxylog 2000 p us Pocket Guide is not a replacement or substitute for the Instructions for Use. Any use of the device requires full understanding and strict observation of the Instructions for Use.
Page 7
Front panel with all options A Screen with screen pages for the specific application K Display symbols for the power supply Charge status of the internal battery B Key Alarms for setting and displaying alarm Mains power supply connected limits L Central rotary knob for making selections / C Key Settings for setting other ventilation...
Side view, right A Emergency air intake B Screw to secure the battery compartment cover C Connectors for flow measuring hoses D Connectors for ventilation hose E Connector for medical gas hose F Socket for DC supply G Window for Infrared Data Association (IrDA) interface...
Page 9
Rear view A Filter cartridge for intake of ambient air B Rating plate...
Preparation Breathing Systems The Oxylog 2000 p us Pocket Guide is not a replacement or substitute for the Instructions for Use. Any use of the device requires full understanding and strict observation of the Instructions for Use. Reusable hose system A Breathing valve B Ventilation hose C Flow and pressure measuring hoses...
Assemble the reusable hose system 1 Place the diaphragm (B) in the breathing valve. Ensure that it is inserted correctly. 2 Fit the cover (A) and turn it approximately 90° clockwise to secure into position. 3 Push the flow sensor (C) into breathing valve. Note the preferred position as indicated by the groove.
Preparation Connect the reusable hose system 1 Connect the ventilation hose (A) to the breathing valve. 2 Connect the flow measuring hoses (B) to the nozzles on the flow sensor. Note the different diameters. 3 Connect the flow measuring (C) hoses to the Oxylog 2000 p us. 4 Connect the ventilation hose (D) to the gas outlet on the Oxylog 2000 p us...
Connect the disposable hose system 1 Connect the blue flow measuring hose (B) to the blue gas connector. 2 Connect the transparent flow measuring hose (A) to the other gas connector. 3 Connect the ventilation hose (C) to the gas outlet on the Oxylog 2000 p us When changing the ventilation hose system If the reusable ventilation hose system is to be used instead of a...
Connecting the power supply Oxylog 2000 p us is designed to operate on power supplies with different voltages: • DC voltage from the on-board power supply: Via DC/DC converter ° With AC/DC power pack ° • With rechargeable battery (specified Smart Battery) Internal supply with rechargeable battery Installing the battery 1 Insert a fully charged battery into the battery compartment (A).
Preparation Charging the battery The green lamp (B ) lights up when the battery is actively charging. A three colored indicator (A) lights up to show the current Ox ylo g 2000 plus charge status of the internal battery: • Green: when the battery has been fully charged. •...
Indication of battery capacity / battery operation The remaining capacity of the battery is indicated by Oxylog 2000 p us in 25% increments in the lower right section of the information window when power is ON. The capacity indication is overwritten if higher priority messages are activated.
Preparation External power supply External power supply from mains power (AC/DC power pack) 1 Connect the mains plug (A) to the mains outlet. 2 Connect the DC plug (B) to the DC connector on the Oxylog 2000 p us O xylog 2000 plus Oxylog 2000 p us 3 When the is connected to an external supply, the...
Page 19
24 V DC (12 V, 24 V, 28 V DC). 28 V DC Oxylog 2000 plus 1 Plug the large connector (A) of the DC/DC converter into the on-board supply. 2 Plug the small connector (B) into the DC connector of the...
Preparation Connecting the gas supply from an O cylinder Connecting the gas supply supply from an O cylinder 1 Use a full O cylinder. 2 Connect the pressure reducer (270 to 600 kPa delivery pressure, 500 kPa nominal pressure) to the O cylinder.
Supply from a piped medical gas system Oxylog 2000 p us 1 Connect the O medical gas hose (A) to the 2 Connect the gas hose (B) to the O terminal unit until the supply of O is confirmed. Preparation...
Preparation Determining the approximate pneumatic operating time for the Oxylog® 2000 p us The pneumatic operation time increases when Oxylog 2000 p us operates with O AirMix, as ambient air is drawn into the device. The amount of gas from the high-pressure supply, which is currently being consumed, is indicated by the Oxylog 2000 p us in the lower left section of the information window in L/min.
Page 23
Determining the approximate pneumatic operating time for the Oxylog® 2000 p us Example for supply of medical gas: • Cylinder pressure measured on the pressure gauge of the pressure reducer: 20,000 kPa (200 bar) • Liquid capacity of the O cylinder: 2.1 L Supply of medical gas: 2.1 L x 20,000 kPa = approximately 420 L...
Operation /Settings Starting operation The Oxylog 2000 p us Pocket Guide is not a replacement or substitute for the Instructions for Use. Any use of the device requires full understanding and strict observation of the Instructions for Use. Switch ON •...
Shutdown After disconnecting the patient switch the ventilator OFF: 1 Press the key (A) for approximately 3 seconds. The yellow lamp flashes and ventilation is terminated by the device. 2 Press the rotary knob (B) to acknowledge the alarm Ox ylo g 2000 plus !!! Confirm device OFF with rotary knob.
Operation /Settings Perform device check Oxylog 2000 plus Connect the test lung Switch ON 2 Press the rotary knob (A) to 1 Connect the angled adapter (A) to 1 To switch ON briefly press confirm, before the bar is full. the breathing valve.
Ventilation controls Selecting the ventilation mode 1. Press the appropriate ventilation mode key (A) for approximately 3 seconds. Oxylog 2000 plus 1 Press the appropriate ventilation mode key (A). 2 Press the rotary knob (B) to confirm. The selected ventilation mode will be activated.
Operation /Settings Preparing ventilation mode Set ventilation parameters 1 Set the required control below the display. 2 Select, set and confirm a parameter on the display with the rotary knob. The former settings are retained if confirmation is not received within 15 seconds.
Operation /Settings VC-CMV / VC-AC VC-CMV – Volume Controlled - Controlled Mandatory Ventilation. Plateau time Volume-controlled ventilation with fixed mandatory minute volume MV, Tplat set with tidal volume VT and respiratory rate RR. plat PEEP Set the ventilation pattern with the controls below the display: •...
Trigger (VC-AC) Activating/setting the trigger 1 Press the key Settings (A) until the trigger parameter is displayed. 2 Select the line Trigger on the display and then set and confirm Ox ylo g 2000 plus the value with the rotary knob. Small value = high sensitivity.
Operation /Settings VC-SIMV (optional PS) VC-SIMV - Volume Controlled - Synchronized Intermittent Mandatory Pressure support PS Ventilation rapid slow For patients with inadequate spontaneous breathing, or for patients who rise time rise time PEEP are to be weaned gradually. Fixed mandatory minute volume MV is set with tidal volume VT and Trigger window ventilation respiratory rate RR.
Page 33
VC-SIMV (optional PS) Pressure support (optional) The following can also be set on the display for VC-SIMV / PS: • Setting on page 1: Pressure support ΔPsupp above PEEP. • Setting on page 2: Pressure rise time Slope flat ramp = long pressure rise time medium ramp = medium pressure rise time steep ramp = short pressure rise time.
Operation /Settings SpnCPAP (optional PS) (optional NIV) SpnCPAP - Continuous Positive Airway Pressure Set the ventilation pattern with the controls below the display: • Maximum airway pressure Pmax. • O setting FiO Pressure support (optional) The following can additionally be set on the display for SpnCPAP / PS: •...
SpnCPAP (Apnea Ventilation) Apnea back-up ventilation is only applicable when using the SpnCPAP Pmax mode. In the event of an apnea, the ventilator will automatically activate volume controlled mandatory ventilation (VC-CMV). Setting apnea ventilation Apnea alarm time Tapn On the display: Start Flow apnea ventilation...
Operation /Settings NIV – Non-Invasive Ventilation (optional) NIV (optional) can only be activated as a supplementary function in the pressure controlled ventilation modes SpnCPAP and SpnCPAP / PS. Mask leakages are detected by the device, compensated and included in the measured values for VTe and MVe. The leakage alarm is inactive. •...
Display operating controls A Values key; to change screen pages in the "Measured Values" window, to display MVe or VTe. B Values key; to change screen pages in the "Measured values" Ox ylo g 2000 plus window, to display the measured values. C Settings key;...
Alarms and Measured Values In the event of an alarm The Oxylog 2000 p us Pocket Guide is not a replacement or substitute for the Instructions for Use. Any use of the device requires full understanding and strict observation of the Instructions for Use. •...
Setting alarm limits To set alarm limits for MV and RRsp • Press the key Alarms (A). Display example Alarms screen with variable alarm limits: = lower alarm limit. Ox ylo g 2000 plus = upper alarm limit. Alarm Range 2 to 41 L/min 0.5 to 40 L/min RRsp...
Alarms and Measured Values Displaying other measured values MVe / VTe window A Parameter measured. B Measured value. C Unit of measure. D Page number. Values window In the values window five different values can be displayed. To switch between the values: •...
Alarm - Cause - Remedy The Oxylog 2000 p us Pocket Guide is not a replacement or substitute for the Instructions for Use. Any use of the device requires full understanding and strict observation of the Instructions for Use. Oxylog 2000 p us classifies alarm messages according to three priority !!! Warning High priority alarm message levels and identifies these accordingly with the aid of exclamation marks: !! Caution...
Trouble Shooting Messages in the alarm window Alarm Cause Remedy !!! Apnea Spontaneous breathing by the patient Ventilate in VC-CMV mode. has failed, or disconnection. Ensure that hose connections are tight. Faulty flow sensor. Replace flow sensor. !!! Apnea ventilation The ventilator has automatically Check ventilation mode.
Page 43
Messages in the alarm window Alarm Cause Remedy !!! Confirm device OFF has been pressed for 3 seconds. To switch OFF: confirm with the rotary knob. with rotary knob To continue ventilation, press key again. !!! Device failure Technical defect. Contact your local DrägerService for additional support.
Page 44
Trouble Shooting Messages in the alarm window Alarm Cause Remedy !! High respiratory rate Patient breathes at a high Check patient's condition, check ventilation spontaneous rate. pattern, correct alarm limit RRsp if necessary. !! Int. battery charging inop Technical defect. Contact your local DrägerService for additional support –...
Page 45
Messages in the alarm window Alarm Cause Remedy !!! Leakage (not in NIV) The measured expiratory tidal volume VTe Repair leaks in hose system and possibly in is approximately 40% lower than the the tube. inspiratory value. Use new flow measuring hoses. Faulty flow sensor.
Page 46
Trouble Shooting Messages in the alarm window Alarm Cause Remedy !!! MV low The minute volume MV has dropped Check patient's condition, check ventilation below its lower alarm limit. pattern, adjust alarm limits if necessary. Leak in exhalation system. Ensure connections in exhalation system are tight.
Page 47
Messages in the alarm window Alarm Cause Remedy !!! Paw high The alarm limit Pmax for the airway pressure Check patient's condition, check has been reached. Patient "fights" the ventilation pattern, adjust alarm ventilator, coughing. limits if necessary. Ventilation hose kinked, or obstructed. Check hose system, breathing valve and tube.
Page 48
Trouble Shooting Messages in the alarm window Alarm Cause Remedy !! Paw measurement inop Fault in flow measurement hoses. Ensure hose system for loose connections. Ensure flow measurement hoses are connected correctly. Technical defect. Contact your local DrägerService for additional support – restricted operation is now possible.
Error messages during the device check Message in the Cause Explanation/Remedy alarm window No communication Device defective. Contact your local DrägerService control- / charge-board System leakage Leak in ventilation hose system and/or test lung. Check hoses, breathing valve, flow sensor and test lung for leaks and replace if necessary.
Trouble Shooting Messages in the information window Message Cause Explanation/Remedy RR = 12 per min or Change in Ti, RR or VT VT = 800 mL in ventilation mode VC-SIMV. I : E = 1 : 1.5 Flow = 15 L/min RR = 12 per min or Change in I:E, RR or VT in VT = 800 mL...
Page 52
REGION EUROPE SOUTH REGION ASIA / PACIFIC HEADQUARTERS Dräger Médical S.A.S. Draeger Medical South East Asia Pte Ltd Drägerwerk AG & Co. KGaA Moislinger Allee 53–55 Parc de Haute Technologie d’Antony 2 25 International Business Park 25, rue Georges Besse #04-27/29 German Centre 23558 Lübeck, Germany 92182 Antony Cedex, France...