Functional Verification: Output Relay Test; Description; Test Procedure; Example Program - Keysight Technologies E1445A Service Manual

Arbitrary function generator, vxibus 75000 series c
Table of Contents

Advertisement

Verification Tests

Functional Verification: Output Relay Test

Description

The purpose of this test is to check the output relay.

Test Procedure

1 Reset the AFG:
2 Set up equipment as shown in Figure 2-4:
Figure 2-4 Output Relay Test Setup
3 Set up the AFG to output a 1 MHz sine wave:
4 Verify that a 1 MHz sine wave appears on the scope.
5 Disable the Output relay:
6 Verify that no signal appears on the scope.

Example Program

This program performs the Functional Verification Tests for the AFG. An Keysight
E1405/E1406 Command Module is required for this test.
28
*RST;*CLS
FREQ 1E6;
:VOLT 4VPP
INIT:IMM
OUTP OFF
Functional Verification: Output Relay Test
Reset AFG and clear status registers
Set freq to 1 MHz
Set AFG amplitude
Initiate
Keysight E1445A Arbitrary Function Generator Service Guide