Re-Assemble The Signal Generator; Verify Signal Generator Calibration - Keysight Technologies E8267D Installation Notes

Psg signal generators upgrade kit for option 601 and 602, digital bus baseband generator with 8 and 64 msample memory, respectively
Hide thumbs Also See for E8267D:
Table of Contents

Advertisement

Activate Option 601/602
1. Follow the instructions on the Entitlement Certificate.
Verify Proper Installation of Option 601/602
1. Turn the signal generator on.
2. Check that Option 601/602 is installed by pressing
>
>
.
Utility
Instrument Info/Help Mode
Options Info
3. Perform the "Checking Signal Generator Functionality" procedure, as described on
page
5.
4. Verify that the Baseband Generator self-tests all pass.

Re-Assemble the Signal Generator

Refer to
Figure
1.
1. Reinstall the inner and outer instrument covers by reversing the order for removal.
2. Torque all T-10 screws to 9 in-lbs.
3. Torque all T-15 and T-20 screws to 21 in-lbs.
4. Attach the Digital Bus label over "DIG I/Q I/O" on the rear panel.
Run the I/Q Calibration
1. Turn the instrument on and let it warm up for 30 minutes.
2. Press:
I/Q
>
I/Q Calibration
>
Calibration Type Full
>
Execute Cal

Verify Signal Generator Calibration

The new baseband generator will function properly in most applications, without re-calibrating the signal generator. If error
vector magnitude (EVM) is critical to the application, the following adjustment/performance tests are recommended. Refer
to the PSG Family Signal Generators Calibration Software:
1. Self Test
2. I/Q Prelevel Gain Mixer Bias, and Digital Gain Cal
3. I/Q Impairment Calibration
4. EVM and IQ Offset
5. External Digital Modulation Level Accuracy Relative to CW
6. Dual Arbitrary Waveform Generator Check
14