Verify Signal Generator Calibration
Verify Signal Generator Calibration
1. Turn the signal generator on and allow it to warm up for 30 minutes.
2. Perform the following adjustment and performance tests, in the order listed, to verify signal
generator calibration. Refer to the PSG Family Signal Generators Service Software.
For units with serial number prefix < US/MY/SG4928:
— Self Tests
— Power Flatness Calibration
— Lowband Attenuator Calibration (<3.2GHz)
— High Band Attenuator Calibration (>3.2GHz)
— V-Band Attenuator Calibration (>50GHz)
— Maximum Leveled Output Power
— Power Level Accuracy
For units with serial number prefix > US/MY/SG4928:
— Self Tests
— High Band Power Flatness Calibration (>3.2GHz)
— Lowband Power Flatness Calibration (<3.2GHz)
— V-Band Power Flatness Calibration (>50GHz)
— Lowband Attenuator Calibration (<3.2GHz)
— High Band Attenuator Calibration (>3.2GHz)
— V-Band Attenuator Calibration (>50GHz)
— Low Power Attenuator Accuracy
— V-Band Low Power Attenuator Accuracy (Opt 567)
Installation Note E8251-90302