Shutdown - Agilent Technologies InfinityLab LC Series User Manual

Variable wavelength detectors
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

7
Error Information
General Error Messages

Shutdown

Error ID: 0063
An external instrument has generated a shutdown signal on the remote line.
The module continually monitors the remote input connectors for status signals.
A LOW signal input on pin 4 of the remote connector generates the error
message.
Probable cause
1
2
3
4
Agilent InfinityLab LC Series VWD User Manual
Leak detected in another module with a CAN
connection to the system.
Leak detected in an external instrument with
a remote connection to the system.
Shut-down in an external instrument with a
remote connection to the system.
The degasser failed to generate sufficient
vacuum for solvent degassing.
Suggested actions
Fix the leak in the external instrument before
restarting the module.
Fix the leak in the external instrument before
restarting the module.
Check external instruments for a shut-down con-
dition.
Check the vacuum degasser for an error condi-
tion. Refer to the Service Manual for the degasser
or the pump that has the degasser built-in.
83