FujiFilm VisualSonics Vevo MD User Manual page 317

Hide thumbs Also See for VisualSonics Vevo MD:
Table of Contents

Advertisement

Did any part of the system or transducer come in contact with
broken skin, blood, or bodily fluids?
(semi-critical)
(non-critical)
Before getting started
Inspect the system and transducer to determine that it is free of any
l
unacceptable deterioration, such as corrosion, discoloration, pitting, or cracked
seals. If damage is evident, discontinue use, and contact FUJIFILM SonoSite
or your local representative.
Confirm that cleaning and disinfecting materials are appropriate for your
l
facility's use. FUJIFILM SonoSite tests disinfectants and disinfectant devices
for use with the system and transducer.
Verify expiration dates, concentration, and efficacy of chemicals (for example,
l
a chemical strip test). Disinfectants and cleaning methods listed are
recommended by FUJIFILM SonoSite for efficacy and material compatibility
with the products.
Cleaning and disinfecting guide
YES
OR
NO
Determining the required cleaning and disinfecting level
"Step A: Semi-critical contact
cleaning and disinfecting of
system and transducer"
page 319
"Step B: Non-critical contact
cleaning and disinfecting of the
system and transducer"
page 326
on
on
317

Advertisement

Table of Contents
loading

Table of Contents