Non-Critical Contact Cleaning And Disinfecting Of The Foot Switch - FujiFilm VisualSonics Vevo MD User Manual

Hide thumbs Also See for VisualSonics Vevo MD:
Table of Contents

Advertisement

Non-critical contact cleaning and disinfecting of the foot switch

Follow the manufacturer's instructions when using cleaners and disinfectants. The
materials listed in the table below are chemically compatible with the foot switch.
Confirm that cleaning and disinfecting materials are appropriate for your facility's use.
Product
Tristel Sporicidal Wipes
Microbac Tissues
Cleanisept wipes
Sani-Cloth AF3 (grey
top)*
Sani-Cloth Bleach
(orange
top)*
Sani-Cloth HB          
(green top)
PI-Spray II
*Qualified for use as an intermediate-level disinfectant for mycobacteria
WARNING:
Disconnect the system from the power before cleaning to avoid
electrical shock.
Wear the appropriate personal protective equipment (PPE) recommended by
the chemical manufacturer, such as eye wear and gloves.
Non-critical contact cleaning and disinfecting of the foot switch
Compatible cleaners and disinfectants
System
(UHF22, UHF48, UHF70)
Transducers
Wet Contact Time
30 seconds
1 minutes
2 minutes
3 minutes
4 minutes
10 minutes
10 minutes
333

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the VisualSonics Vevo MD and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents