16 Obtain a new cloth moistened with a compatible disinfectant or disinfectant wipe.
Compatible disinfectant materials are listed in the table
"Compatible cleaners and
disinfectants"
on page 326.
17 Disinfect the transducer body by wiping the surface.
18 Ensure the transducer stays wet with the disinfectant for the wet contact time
listed in the table
"Compatible cleaners and disinfectants"
on page 326. This may
require reapplication to achieve the required wet time.
♦ For example, if you are using Sani-Cloth Bleach it was a wet contact time of 4
minutes. Wipe the transducer cable and body until thoroughly wet. If the transducer
starts to dry before 4 minutes is reached, use a wipe and re-wet the transducer.
Continue the process of re-wetting with a new wipe until a total wet time of 4 minutes
is reached.
19 Allow the transducer to air dry in a clean, well ventilated space.
20 Examine the transducer and cable for damage, such as cracks or splitting where
fluid can enter. If damage is evident, do not use the transducer, and contact
FUJIFILM SonoSite or your local representative.