FujiFilm VisualSonics Vevo MD User Manual page 382

Hide thumbs Also See for VisualSonics Vevo MD:
Table of Contents

Advertisement

MI and TI output display accuracy
The accuracy result for the MI is stated statistically. With 95% confidence, 95% of the
measured MI values will be within +61.0% to -73.1% of the displayed MI value, or
+0.2 of the displayed value, whichever value is larger.
The accuracy result for the TI is stated statistically. With 95% confidence, 95% of the
measured TI values will be within +118.8% to -144% of the displayed TI value, or +0.2
of the displayed value, whichever value is larger.
A displayed value of 0.0 for MI or TI means that the calculated estimate for the index
is less than 0.05.
382
Appendices