FujiFilm VisualSonics Vevo MD User Manual page 322

Hide thumbs Also See for VisualSonics Vevo MD:
Table of Contents

Advertisement

10 Clean the system to remove any visible gel, debris or bodily fluids.
♦ Wipe from clean areas to the soiled areas to avoid cross-contamination.
11 Verify that all gel, debris, and bodily fluids have been removed from the transducer
and system. If necessary, repeat Steps 9 and 10, with a new wipe.
12 Obtain a compatible disinfectant. Compatible disinfectant materials listed in the
table
"Compatible cleaners and disinfectants"
Caution:
Make sure the solution does not get into any of the connector ports.
Apply the disinfectant solution to the cloth rather than applying it directly to
surfaces if not using a wipe.
13 Disinfect the system by wiping all surfaces including:
clinical display (upper monitor)
l
control panel (lower monitor)
l
transducer and gel holders
l
cable hooks
l
grab bars
l
all surfaces that may have been touched and/or contaminated
l
14 Ensure the system stays wet with the disinfectant for the wet contact time listed
in the table
"Compatible cleaners and disinfectants"
reapplication to achieve the required wet time.
♦ For example, if you are using Sani-Cloth Bleach it has a wet contact time of 4
minutes. Wipe the entire system until thoroughly wet. If the system starts to dry
before the 4 minutes is reached, use a new wipe and re-wet the system. Continue the
process of re-wetting with a new wipe until a total wet time of 4 minutes is reached.
15 Allow the system to air dry in a clean, well ventilated space. Do not dry with a
cloth.
322
Cleaning and disinfecting
on page 319.
on page 319. This may require

Advertisement

Table of Contents
loading

Table of Contents