4 In the Location list, select the location of the system.
5 In the list of devices, select the check box next to one or more archivers, or worklist
servers.
You can select a maximum of four archivers, and one worklist server for each
location. Only one archiver can be associated with an MPPS and storage
commitment server. Selected devices have a check mark in the left-hand column.
6 (Optional) If you want to use the MPPS service, associate the MPPS server with
the archiver:
Note:
Only one MPPS server can be selected for a location.
i. Select the check box of the MPPS server that you want (MPPS servers appear
near the end of the list)
ii. Select the check box of the archiver
iii. Select the check box in the archiver's MPPS column
7 (Optional) If you want to use the storage commitment service, associate the
storage commitment server with the archiver:
Note:
Only one storage commitment server can be selected for a location.
i. Select the check box for the storage commitment server that you want (storage
commitment servers appear at the end of the list)
ii. Select the check box for the archiver
iii. Select the check box in the archiver's storage commit (SC) column
8 Complete any additional configuration tasks, and then tap Done.
Associating devices with locations
91