Problem
Unable to create new
exams
Unable to archive an
exam
Problem
Lack of penetration or
sensitivity
Problem
Color bands in the
image
Respiration artifacts in
the image
412
Appendices
Patient management troubleshooting
Ensure that a transducer is connected to the front
l
panel of the Vevo MD Imaging System, and ensure
that it has been initialized.
Ensure that the Ethernet cable is attached and that
l
the connection is active.
2D and M-Mode troubleshooting
Ensure that there is adequate coupling medium (for
l
example, ultrasound gel) between the transducer
and the patient.
Adjust the position of the TGC sliders.
l
Ensure the appropriate transducer is being used.
l
Color Doppler Mode troubleshooting
Adjust wall filter setting.
l
Adjust the priority setting.
l
Adjust line density setting.
l
Adjust wall filter setting.
l
Solution
Solution
Solution