Tap Presets to see the available presets and then select an appropriate set of
optimized image acquisition settings. This button is active during image acquisition in
every mode.
The list of presets includes the factory default presets as well as any custom presets
you have added to the system. All presets are mode, exam type, transducer, and user
dependent.
Adjusts the priority relationship between the overlay data and the background 2D data
so you can eliminate false readings. Priority determines the threshold point on the
gray scale above which the system does not apply color data. The red marker along
the left side of the display map indicates the threshold point.
Tap the Priority button and use the Up and Down buttons to select the required
value.
262