Problem
Transducer is not
recognized by the
system
Patient data is
corrupted
Follow the steps; write down the error information,
l
call FUJIFILM SonoSite customer support, and
power off the system.
Disconnect the transducer, then reconnect the
l
transducer.
Clean the transducer interface on the system using a
l
compressed air duster—never use a cloth or wipes
as the pins on the connector could be damaged by
direct contact.
Patient data will not be available to be review,
l
archive or export.
When exporting or archiving patient data, ensure that
l
the USB device or Ethernet cable you are using, is
not removed before the process is complete.
Solution
Troubleshooting
411